Bresle Elcometer 138 salt contamination meter
Equipped with an integrated flat sensor, the Bresle 138 Salt Contamination Meter measures the conductivity of a sample, automatically converts it and displays the salt density. This eliminates the need for operators to perform calculations themselves to work in compliance with ISO 8502-6 or ISO 8502-9.
Key features of the Bresle Elcometer 138 Salt Contamination Meter :
Possibility of obtaining highly accurate measurements with a single drop
Out of range" and "low battery" indicators
Visual indicator when the ambient temperature is outside the working range
Spécifications
Measurement principle
2 AC bipolar electrodes
Measurement mode
ISO, IMO, Temperature
Minimum sample volume
0,12ml
Measurement range
Mode ISO
0 - 2399µg/cm²
Mode IMO
0 - 2199µg/cm²
Conversion factor
Mode ISO
µS/cm à µg/cm² : 0,12
µS/cm à mg/m² : 1,2
Mode IMO
µS/cm à µg/cm² : 0,11
µS/cm à mg/m² : 1,1
Resolution
Mode ISO
0 - 239,9µg/cm² : 0,1µg/cm²
240 - 2399µg/cm² : 1µg/cm²
Mode IMO
0 - 219,9µg/cm² : 0,1µg/cm²
220 - 2199µg/cm² : 1µg/cm²
Precision
±2% full scale (for each range)
Operating temperature
5°C à 40°C (41°F à 104°F)
Operating humidity
Relative humidity of 85% or less (no condensation)
Battery type
2 x CR2032 lithium batteries
Autonomy
approx. 200 hours continuous use without backlighting
Weight
50 g (1.76oz) - with sensor and batteries
Dimensions
164 x 29 x 20mm (6,5 x 1,2 x 0,8”)
Packing list
Elcometer 138 Bresle Salt Contamination Meter, 14ml (0.5fl oz) bottle of conditioning solution, 2 x CR2032 lithium batteries and instructions for use